System specification

MEG-KIT system description

MEG systems are equipped with highly sensitive sensors called SQUIDs. For SQUIDs to operate and become sensitive to very small changes in magnetic field, they must be brought to a superconducting state, in which the electrical resistance of the sensor material vanishes. To reach this state, the sensors are cooled down to -269 degrees Celsius with liquid helium, in which they are constantly bathed. A heater system ensures that the liquid helium stays in a liquid state.

The MEG-KIT system has:

  • 208 axial gradiometers used to measure brain activity

  • 8 orthogonally-oriented (reference) magnetometers located in the dewar but away from the brain area, used to measure and reduce external noise offline

  • 32 open positions, of which we currently use 8 to record stimulus triggers

  • 24 channels to record Eye Tracker data directly, auditory signals from our mixer and vocalization information from our optoacoustic fiber-optic microphone

Warning

The NYUAD MEG system uses liquid helium to keep the SQUID sensors cold (-269°C, -452°F). If the helium tank leaks, the liquid helium immediately boils off, which can cause rapid suffocation and severe frostbite. Handle the system with care and, in particular, never subject the dewar to shocks. If you notice anything unusual during an experiment (for example, you suspect a helium leak), open the MSR and release the participant immediately. Do not leave the room while a participant is in the MSR.

Important

Since 2018, the gradiometer sensor mapped to channel 091 (KIT indexing) has been irreparably damaged. Since then, the sensor has been statically typed as a magnetometer, which ensures that its data is not processed automatically by the provided analysis pipelines. During a data acquisition, the channel displays a square TTL signal.

Lab setup

Lab setup: Outside the MSR

Computers

  • MEG DAQ PC: used to acquire the MEG data. It has four screens for visualising the sensor data in real time.

  • Stimulus1 PC: used to run the experiments. It has two screens: the bottom screen shows the code, and the top screen shows the experiment, mirroring the image projected inside the MSR.

Vpixx stimulus/projector system

The Vpixx system presents the visual stimuli and collects the participant’s responses. It consists of the Propixx projector and the Responsepixx button boxes described below.

Propixx (Screen projector)

The Vpixx Propixx system projects visual stimuli onto a screen that is 65 cm wide.

Responsepixx boxes (Button boxes)

The boxes currently used in the MEG lab are listed below.

Responsepixx boxes

ResponseBox serial number

Type (left or right)

White

Red

Yellow

Green

Blue

22001-1049

Left

1001000000

0001100000

STUCK

0011000000

0101000000

22002-1071

Right

0001010000

0001000001

0001000010

0001000100

0001001000

The boxes provide 10 buttons in total, coded on 10 bits (bit 0 to bit 9).

Current Designs fORP dial and 905 interface

In addition to the Responsepixx boxes, the lab has a fiber-optic rotary dial from Current Designs (fORP system), for rating, tracking and adjustment tasks. How to read it from Psychtoolbox, with ready-to-use functions and a rating-task example, is on the dial experiment page.

fORP dial hardware

Item

Model (SKU)

Details

Dial (response device)

HHSC-1x1-DIAL

All plastic, no metal or electronics, so it is MR/MEG safe and is used inside the MSR. Circular dial (4.8 cm diameter) turned left or right, nominally 1 degree per step (360 steps per turn), 3 m (10 ft) fiber pigtail.

Interface unit

FIU-905

Converts the optical signals from the dial into USB keyboard taps for the stimulus PC. Contains metal and electronics: keep it outside the MSR.

Because the dial pigtail is only 3 m long, a Current Designs fiber bundle extension is used through the MSR waveguide between the dial and the 905.

FIU-905 ports (from the 905 Getting Started guide in the Current Designs Tech Docs, which need a customer login at https://www.curdes.com/docs/)

  • Front panel

    • P1, mini-USB (5-pin): the only data output, and it also powers the unit (no power supply). Connect it to the stimulus PC. The 905 appears as a standard USB HID device (keyboard + mouse + joystick interfaces, named “904”); no driver needed.

    • 6 LEDs: TRIGGER, B1, B2, B3, B4, B5. With the dial, TRIGGER flashes on every clockwise step and B1 on every counter-clockwise step; B2 to B5 stay off. Both blinking continuously while nobody touches the dial is rest chatter (see below), not a fault.

  • Rear panel

    • P4, optical port: the fiber bundle extension, or the dial pigtail directly.

    • P3, TTL trigger input: gold SMB coaxial connector, positive edge triggered, weak 47 kOhm internal pull-up, no minimum pulse width. Use the CB-SMB-BNC-1 cable for a BNC source.

    • P2, optical trigger input (HFBR type): for optical triggers from MRI scanners. Not used on the KIT MEG.

What the 905 sends

  • Button boxes: buttons 1 to 5 send the keys 1 to 5 (6 to 0 on a unit configured as LEFT); a TTL or optical trigger sends t.

  • Dial: the 905 recognises the dial and decodes it itself. Every step is one short (about 1 ms) key tap: t for one step clockwise, b for one step counter-clockwise. The position is the number of t taps minus the number of b taps.

  • Rest chatter: when the dial rests exactly on an encoder edge, the 905 may send alternating t b t b ... taps at more than 100 per second (TRIGGER and B1 LEDs blink). They cancel out in the t - b count, but they are typed into whatever window has focus, so unplug the 905 USB when it is not in use.

Connecting the dial and setting the gain

The 905 sets its optical gain at the moment its USB cable is plugged in, from the light it sees at that moment. For a button box that is enough. For the dial, if the dial is still at that moment the gain is wrong for a rotary encoder and the 905 mis-decodes the direction (one turn gives about as many b as t, in a repeating pattern). Therefore:

  1. Dial inside the MSR, fiber bundle through the waveguide, 905 outside the MSR next to the stimulus PC.

  2. Connect the dial (and the bundle) to P4 before the USB cable.

  3. Have someone turn the dial continuously, plug in the 905 USB cable while the dial is turning, and keep turning for a few seconds.

  4. Check with dial_dot_demo (see the dial experiment page): one full turn of the dial should move the dot once round the circle, in the same direction.

This procedure has to be repeated whenever the 905 loses USB power (cable unplugged, PC cold boot). It does not need repeating between MATLAB sessions.

Note

Measured on 2026-11-04 with the bundle in the path: gain set with the dial at rest gave 5 taps per encoder cycle with a net of only 1 step (about 10% decode quality); gain set with the dial turning gave runs of 300 consecutive same-letter taps per turn. The Current Designs product page says the 905 does not support pointing devices, but the dial page lists the 905 as compatible, and with the procedure above it works. Support: support@curdes.com, +1 215-387-5456.

Lab setup: Inside the MSR

The MSR is equipped with six dimmable halogen light bulbs; replacement bulbs are kept in the lab. The backup battery is located in one of the office spaces and is connected via yellow cables.

  • Lights: the switch is located by the door.

  • Dewar: contains the liquid helium in which the SQUID sensors, spread across the helmet, are submerged.

  • Foam insert headphones: deliver the audio stimuli to the participant, as well as spoken instructions from the operator.

  • Marker box: connects the head position indicator coils (marked with 5 colours).

  • Heater cable: keeps frost and ice from forming on the dewar; the switch is located at the back of the MSR.

  • Projector: back-projects the stimulus onto a screen; the remote control is outside the room.

  • Amplitude switches: control the signal strength of the marker coils.

Light button setup Eyetracker and MEG Helmet KIT amplitude switches

Electronic lag times between visual stimulus and KIT trigger box

A trigger signal marking a visual stimulus, sent from the Vpixx system to the KIT DAQ computers, arrives slightly before the visual stimulus is displayed by the Vpixx projector, because of delays in the electronic circuits. This was investigated in detail by Gayathri Satheesh with a photodiode experiment of 1000 trials sampled at 1000 Hz. Her findings are summarised as follows:

  • the mean lag is -8.27852 ms (the negative sign means that the trigger signal appears in the data 8.27 ms before the participant sees the visual stimulus)

  • the standard deviation of the lag is 0.534153 ms

  • the maximum lag was 11 ms

  • the minimum lag was 8 ms

Therefore, when processing MEG data that requires millisecond precision, correct the visual stimulus trigger signals by shifting them 8.2 ms later in time.

This was later confirmed by a counterbalanced two-recording photodiode measurement (trigger to projector +8.36 ms, sd 0.48; trigger to stimulus monitor +2.05 ms; stimulus monitor to projector +6.31 ms). See Display and trigger timing for the full method, correction code and acquisition guidance.

MEG-Channels

Note

Channel numbering in the KIT data acquisition software (MEG160) starts at index 0, whereas in MATLAB channels start at index 1.

MEG-KIT sensor channels

  • Channels 0 to 90 and 92 to 207: SQUID gradiometers

  • Channel 91: broken gradiometer sensor, typed as a reference magnetometer so that its data is not processed automatically by the analysis pipelines

  • Channels 208 to 223: reference magnetometers (used for denoising and for characterising the ambient magnetic field of the environment)

Vpixx Stimulus channels

These channels are free for the experiment designer to use; together they form an 8-bit binary signal.

  • 224: Event marker bit 0

  • 225: Event marker bit 1

  • 226: Event marker bit 2

  • 227: Event marker bit 3

  • 228: Event marker bit 4

  • 229: Event marker bit 5

  • 230: Event marker bit 6

  • 231: Event marker bit 7

Photodiode channel

The photodiode is connected to:

  • 232: Photodiode output signal

Eyetracking channels

The SR Research Eyetracker is connected to analog channels of the KIT-MEG system:

  • 233: eyetracker sample onset linked to Eyetracker Analog Card pin ST

  • 234: x-position of the eye linked to Eyetracker Analog Card pin DAC0

  • 235: y-position of the eye linked to Eyetracker Analog Card pin DAC1

  • 236: linked to Eyetracker Analog Card pin DAC2

  • 237: linked to Eyetracker Analog Card pin DAC3

  • 238: linked to Eyetracker Analog Card pin DAC4

  • 239: linked to Eyetracker Analog Card pin DAC5

Analog Data Output Assignments

Eye tracking mode

Analog output mapping

Channels available

DAC0 (MEG-KIT channel 234)

DAC1 (MEG-KIT channel 235)

DAC2 (MEG-KIT channel 236)

DAC3 (MEG-KIT channel 237)

DAC4 (MEG-KIT channel 238)

DAC5 (MEG-KIT channel 239)

left / right

Monocular

6

X

Y

P

--

--

--

Binocular

Monocular

6

left X

left Y

left P

right X

right Y

right P

Left

Binocular

6

left X

left Y

left P

--

--

--

Right

Binocular

6

--

--

--

right X

right Y

right P

Binocular

Binocular

6

left X

left Y

left P

right X

right Y

right P

left / right

Monocular

4

X

Y

P

--

--

--

Binocular

Monocular

4

left X

left Y

right X

right Y

--

--

Left

Binocular

4

left X

left Y

--

--

--

--

Right

Binocular

4

--

--

right X

right Y

--

--

Binocular

Binocular

4

left X

left Y

right X

right Y

--

--

To enable the analog data output, add the following line to your Psychtoolbox experiment after opening the connection to the eye tracker:

Eyelink('command','analog_out_data_type = GAZE');

MEG-Racks

The KIT-MEG system has 7 racks.

MEG Racks System

MSR: Magnetically Shielded Room

The KIT-MEG is located in an MSR built by Vacuumschmelze.

The detailed MSR schematics are available in the internal documentation.

MSR schematic 1 MSR schematic 2

Data acquisition (DAQ) units

Data acquisition interface (DAQ) of KIT MEG
  • The DAQ consists of four PCs working together, interfacing with MEG160 during data acquisition.

  • The DAQ units are located in a cabinet labelled “DAQ”.

    • The units are labelled ‘DAQ0’, ‘DAQ1’, ‘DAQ2’ and ‘DAQ3’.

    • The system battery of DAQ0 was last changed on February 4, 2020.

    • After a battery change, the BIOS settings might need to be updated. To enter the settings page, press F2 while the system is booting.

      • Make sure Power Management is Enabled.

      • Make sure Low Power Mode is Disabled.

    • If one or more of the DAQ units will not boot, try turning it on manually by pressing the power button on the unit.

    • Inform other lab members if this happens; chances are they have encountered the same or a similar problem.

MEG160 Software parameters

MEG160 is the main software for data acquisition from the KIT-MEG system.

Auto-tuning should be performed before any data acquisition. In the software, open Acquire Measurement. A sensor check can be performed from this window: the circuit of a sensor is working correctly when a sinusoidal wave is seen on the channel corresponding to that sensor.

The following table shows the default parameters that should be set in the acquisition window.

MEG160 parameters

Parameter

Description

Value

Value description

HPF

High Pass Filter

0.1Hz

Let pass only above 0.1 Hz (because electronic circuits generate a small noise below 0.1Hz)

LPF

Low Pass Filter

500Hz

Let pass only below 500 Hz

BEF

Band Elimination Filter

THRU

Let pass all frequencies (BEF turned off)

If strong noise is present at a specific frequency (for example 50 Hz), use the band-elimination filter (BEF) at that frequency to remove it.

Frequently asked questions (FAQ)

Can opening the MSR door while the SQUID sensors are locked damage the SQUIDs?

Opening the magnetically shielded room (MSR) door while the SQUID sensors are locked may cause the output of some channels to appear flat. However, this does not indicate permanent damage to the sensors. The temporary signal suppression is typically due to large fluctuations of the external magnetic field penetrating the MSR. The sensors are not malfunctioning; rather, they momentarily lose their proper operating conditions because of the external magnetic disturbance. Once the MSR door is closed, unlocking and relocking the SQUIDs and then running the auto-tuning restores normal sensor function.

How can I test the trigger channels with the MSR door open?

In this scenario, do not lock the sensors. Keep the MSR door open and run all the tests of your experiment with the Vpixx response boxes, the microphone and the other hardware.

Contact

Name

Email

Number

Role

Hadi Zaatiti

hz3752@nyu.edu

+971 56 275 4921

Research Scientist

Osama Abdullah

osama.abdullah@nyu.edu

NA

Senior Scientist

Yoshiaki Adachi

adachi@ael.kanazawa-it.ac.jp

NA

MEG-KIT machine constructor reference

Jun Kawai

j-kawai@neptune.kanazawa-it.ac.jp

NA

MEG-KIT machine constructor reference

References

The following references are useful for a deeper understanding of MEG systems.