System specification
System Overview
The OPM system was installed on 4 March 2024.
Fieldline OPM system with 96 sensors, installed at NYUAD.
Powering on or off the OPM
The OPM racks must always be turned on or off starting from the top rack.
Sensor locations on helmet
The sensor locations on the helmet are shown in the image /usr/share/hedscan/doc/Beta2SensorLocations.png on the OPM computer.
In addition, every saved .fif file contains the location of the channels in (x, y, z) coordinates.
Trigger channels on OPM
OPM to Vpixx digital pins connection
A custom cable has been built to link the Vpixx system to the OPM system. The pin mapping is given below; the cable datasheet is available in the internal version of this page.
Pin |
Sender: DB25 Digital Output (female port) VPIXX |
Receiver: DB37 Digital Input (male port) OPM |
|---|---|---|
1 |
Dout0 |
Data Bit 0 (pin2) |
2 |
Dout2 |
Data Bit 2 (pin4) |
3 |
Dout4 |
Data Bit 4 (pin6) |
4 |
Dout6 |
Data Bit 6 (pin8) |
5 |
Dout8 |
Data Bit 8 (pin11) |
6 |
Dout10 |
Data Bit 10 (pin13) |
7 |
Dout12 |
Data Bit 12 (pin15) |
8 |
Dout14 |
Data Bit 14 (pin17) |
9 |
Dout16 |
Data Bit 16 (pin20) |
10 |
Dout18 |
Data Bit 18 (pin22) |
11 |
Dout20 |
Data Bit 20 (pin24) |
12 |
Dout22 |
Data Bit 22 (pin26) |
13 |
Ground |
Ground (pin1) |
14 |
Dout1 |
Data Bit 1 (pin3) |
15 |
Dout3 |
Data Bit 3 (pin5) |
16 |
Dout5 |
Data Bit 5 (pin7) |
17 |
Dout7 |
Data Bit 7 (pin9) |
18 |
Dout9 |
Data Bit 9 (pin12) |
19 |
Dout11 |
Data Bit 11 (pin14) |
20 |
Dout13 |
Data Bit 13 (pin16) |
21 |
Dout15 |
Data Bit 15 (pin18) |
22 |
Dout17 |
Data Bit 17 (pin21) |
23 |
Dout19 |
Data Bit 19 (pin23) |
24 |
Dout21 |
Data Bit 21 (pin25) |
25 |
Dout23 |
Data Bit 23 (pin27) |
OPM sensors and direction of measurement
In its current state, the OPM system measures the radial magnetic field (Bz) but not yet the tangential directions (Bx or By). The hardware is capable of measuring both, but FieldLine has not yet released the update that enables the tangential directions to be measured.
Dynamic Range
The dynamic range of a sensor is the ratio between the highest amplitude it can measure linearly, before reaching saturation, and its noise floor, below which a signal cannot be distinguished from noise.
It is often expressed in decibels (dB) as
To compute the dynamic range of an MEG sensor:
measure the highest amplitude \(B_\text{max}\) that the sensor can measure, for example by applying a slowly varying (DC) source and observing the measurement
measure the noise level \(B_\text{noise}\), for example by recording empty-room data and computing the average level of the noise (a measured signal can be distinguished from the sensor noise only when its amplitude is above this level)
OPM HPI Coils
The HPI coils are placed on specific landmarks on the head of the participant.
The 6 HPI coils are connected from the OPM system into the MSR through 6 brass cables with the following characteristics:
TNP 3.5mm Mono Extension (25FT) - 12V Trigger
IR Infrared Sensor Receiver Extension Extender
3.5mm 1/8” TS Monaural Mini Mono Audio Plug Jack Connector Male to Female Cable Wire Cord
Length: 7.6 meters
A script must be run before starting the experiment to activate the HPI coils. The script energises the HPI coils with known sinusoidal waves (see Appendix D of the OPM documentation).
The script can be run at the beginning and at the end of the experiment to register the coil positions, which are then averaged. Coregistration strategies are discussed in the Operational Protocol.